Label: LEUCOFLEX CGP- anticoagulant citrate phosphate dextrose solution with additive solution as-1 kit
- NDC Code(s): 14498-002-01
- Packager: Maco Productions
- Category: HUMAN PRESCRIPTION DRUG LABEL
- DEA Schedule: None
- Marketing Status: New Drug Application
Drug Label Information
Updated December 5, 2019
If you are a consumer or patient please visit this version.
- Download DRUG LABEL INFO: PDF XML
- Official Label (Printer Friendly)
-
DescriptionThe Leucoflex CGP product consists of a blood collection system with 70ml of CPD for the collection of 500±50ml of blood and 100ml of AS-1 Additive Solution. They are supplied with sterile ...
-
Intended UseThe "Leucoflex CGP Leukocyte Reduction Filter" is intended for the leukocyte reduction of Red Blood Cells prepared from whole blood stored at 1-6°C up to 24 hours after collection. The collection ...
-
WARNINGAvoid contact with sharp objects. DO NOT USE if the overwrap or blood bag system shows any signs of deterioration. DO NOT USE if the solutions are not clear. Dispose of all system components that ...
-
CAUTIONSDo not fold or squeeze the Leucoflex CGP filter. Inappropriate handling may adversely affect filtration. Check for kinks in the tubing prior to collection and filtration. Rx only. The ...
-
COLLECTION AND SAMPLING PROCEDUREINSTRUCTION FOR USE - Before Donation - 1) Peel open the transparent overwrap (D) using the corner flap. 2) Support blood bag system as far as possible below donor arm, ideally on an automated ...
-
COMPONENT PREPARATION GENERALCAUTION: Do not fold or manhandle the filters (L). Do not apply pressure, either mechanical or human, to increase the flow rate of the filter. Units with decreased or extended filtration ...
-
BLOOD COMPONENTS SEPARATION1) Obtain collected Whole Blood from 1-6 C storage container. 2) Mix the Whole Blood thoroughly. 3) Pack the collection system into a centrifuge as per your facility's standard operating ...
-
RED BLOOD CELLS LEUKOCYTES REDUCTION1) Lift the additive solution (AS-1) bag above the RBC bag on the filtration rack or drip stand. 2) Break the break-away cannula on the RBC bag by bending the cannula with side-to-side ...
-
HANDLING OF PORT PROTECTORS1) Using both hands tear the port protector as indicated. (O) 2) Hold down the port protector pieces with one hand. (P) 3) Using the free hand, take the spike and engage intor the port. (Q) 4 ...
-
LABELSBag 1 Anticoagulant Citrate Phosphate Dextrose Solution, USP; Bag 2 Additive Solution AS-1; Overwrap; Carton - Anticoagulant Citrate Phosphate Dextrose Solution, USP - Bag 1 Label - Additive ...
-
INGREDIENTS AND APPEARANCEProduct Information