CPDA-1 - anticoagulant citrate phosphate dextrose adenine solution
Fenwal, Inc.
----------
4R3622, 4R3626
FRESENIUS KABI
Fenwal Blood-Pack Units
Rx only
Using Anticoagulant Citrate Phosphate Dextrose Adenine Solution, USP (CPDA-1) and Fenwal HighFlo Needle
Instructions for Use
Collection Procedure:
Use aseptic technique
Note: Nominal tubing dimensions of product are 0.118” inner diameter x 0.025” wall thickness.
Note: If the Y-Sampling Site is not used, donor samples may be collected using an alternate method following standard procedures.
Precautions:
• Do not use unless the solution is clear.
• Before beginning procedure, obtain one access device for each Blood-Pack unit with Y-Sampling Site to be processed.
1. Identify Blood-Pack unit using appropriate donor identification system.
2. Adjust donor scale to desired collection weight and position primary container on the donor scale as far as possible below donor arm.
3. Clamp donor tubing between Fenwal HighFlo1 needle and primary container. This step may be performed prior to step 1 or 2.
4. Following blood center procedures, apply pressure to donor’s arm and disinfect site of venipuncture.
5. Remove needle cover per instructions below:
a) Holding the hub and cover near the tamper-evident seal, twist cover and hub in opposite directions to break seal.
b) Remove needle cover, being careful not to drag the cover across the needle point.
6. Following blood center procedures, perform venipuncture, appropriately secure donor needle and/or tubing and release clamp on donor tubing.
7. When good blood flow is established, stabilize the front of the needle guard to arm with tape. (see Figure 1)
8. Mix blood and anticoagulant in primary container at several intervals during collection and immediately after collection.
9. Collect the appropriate volume based on Blood-Pack unit used.
Note: The volume of anticoagulant is sufficient for the blood collection indicated on Blood-Pack unit ± 10%.
10. Release the pressure on the donor’s arm as appropriate.
Precaution: Do not proceed with the remaining steps until the entire whole blood unit is collected.
11. To avoid possible contamination of the whole blood unit, before filling whole blood sample tubes, hermetically seal the donor tubing near the Y-Sampling Site on the side leading to the primary container using a metal clip or appropriate alternate method.
Precaution: Complete steps 12 - 20 within approximately 4 minutes after sealing the donor tubing to avoid possible clot formation in the tubing.
12. To collect samples, insert the access device by pushing firmly into the Y-Sampling Site until the membrane seal is penetrated (see Figure 2).
Note: If the access device is assembled such that the outer barrel is screwed onto the Luer, make sure to rotate clockwise upon insertion to avoid barrel detaching from Luer.
13. Open the cap on the access device (if applicable).
14. Directly align the vacuum sample tube with the internal needle in the access device. Insert vacuum sample tube into device until the stopper is punctured.
15. Allow vacuum sample tube to fill with blood then remove from the access device.
16. Repeat steps 14 and 15 until the desired number of vacuum sample tubes have been filled.
Notes:
• If the access device needs to be replaced, clamp the tubing between the needle and the Y-Sampling Site. Then, grasp base of Sampling Site with one hand and pull the access device out with the other hand. Firmly insert the new access device. Remove clamp and continue sampling.
• If the access device is assembled such that the outer barrel is screwed onto the Luer, make sure to rotate clockwise upon removal to avoid barrel detaching from Luer.
• The access device can only be replaced one time.
Precaution: When replacing the access device, be careful to avoid contact with any blood droplets on the Luer or Sampling Site. Discard used access device appropriately.
17. Release remaining pressure on donor’s arm.
18. If desired, apply clamp to donor tubing between needle and Y-Sampling Site.
19. Withdrawal of Needle (see Figure 3)
Precaution: The needle guard must be held stationary while the needle is withdrawn into it.
a) Place folded sterile gauze over puncture site and hold in place with finger tip without exerting pressure.
b) Hold sides of the needle guard near the front, between the index finger and thumb. Pull the hub back smoothly until the needle is completely enclosed and securely locked into the needle guard.
c) Confirm the needle is completely enclosed and securely locked into the needle guard.
20. Strip blood from donor tubing into primary container, mix and allow the tubing to refill; repeat once.
21. Seal at X marks on donor tubing to provide numbered aliquots of anticoagulated blood for typing or crossmatching.
Note: Step 22 may be performed prior to step 20 or 21 if desired.
22. Remove and discard the Y-Sampling Site and the donor needle in the needle guard into an appropriate biohazardous waste container following established procedures.
23. Component Preparation:
• If a platelet concentrate is to be prepared, it should be separated from the red blood cells within 8 hours after blood collection.
• Fresh frozen plasma should be separated from the red blood cells and placed in the freezer at -18°C or colder within 8 hours after blood collection.
24. At the appropriate time, prepare the Blood-Pack unit for centrifugation by thoroughly mixing the primary container end over end, then load the unit in a centrifuge cup per the instructions on page 3.
25. Following centrifugation, remove containers from the centrifugation cup taking care not to disturb the red blood cell / plasma interface.
26. Place primary container in plasma extractor and express plasma into the appropriate empty Transfer Pack container by releasing pressure plate and opening closure in tubing of primary container.
27. When desired amount of plasma has been removed, clamp tubing between Y and plasma container.
28. Hermetically seal and separate transfer tubing between the Y-connector and primary container. Be careful to avoid fluid splatter.
29. For further processing with multiple Blood-Pack units, use standard component processing and storage techniques.
30. Store suspended CPDA-1 Whole Blood/Red Blood Cells between 1 and 6°C.
31. Infuse CPDA-1 Whole Blood/Red Blood Cells within 35 days of collection.
Store at Controlled Room Temperature. Protect from freezing. Avoid excessive heat. Definition of “Controlled Room Temperature”:
“A temperature maintained thermostatically that encompasses the usual and customary working environment of 20° to 25°C (68° to 77°F); that results in a mean kinetic temperature calculated to be not more than 25°C; and that allows for excursions between 15°C and 30°C (59° and 86°F) that are experienced in pharmacies, hospitals, and warehouses. Provided the mean kinetic temperature remains in the allowed range, transient spikes up to 40°C are permitted as long as they do not exceed 24 hours ...
The mean kinetic temperature is a calculated value that may be used as an isothermal storage temperature that simulates the non isothermal effects of storage temperature variations.”
Reference: United States Pharmacopeia, General Notices. United States Pharmacopeial Convention, Inc. 12601 Twinbrook Parkway, Rockville, MD.
1 Van der Meer, P.F., & de Korte, D. “Increase of blood donation speed by optimizing the needle-to-tubing connection: an application of donation software.” Vox Sanguinis 2009, 97: 21-25
Code 4R3626
24 Units
Fenwal Blood-Pack Units
Double
Anticoagulant Citrate Phosphate Dextrose Adenine Solution, USP (CPDA-1)
For Collection and Processing of 500 mL Blood
Y-Sampling Site 16 ga. Ultra Thin Wall Fenwal HighFlo Needle
Rx only
Each unit consists of a primary container with 70 mL of CPDA-1 solution containing 2.23 g Dextrose (monohydrate) USP, 1.84 g Sodium Citrate (dihydrate) USP, 209 mg Citric Acid (anhydrous) USP, 155 mg Monobasic Sodium Phosphate (monohydrate) USP and 19.3 mg Adenine USP, pH may have been adjusted with sodium hydroxide; one empty 400 mL Transfer Pack container.
Sterile, non-pyrogenic fluid path.
See instructions for use.
Single use only.
Store at Controlled Room Temperature (refer to direction insert). Protect from freezing. Avoid excessive heat.
Manufacturer
Fresenius Kabi AG.
61346 Bad Homburg / Germany
www.fresenius-kabi.com
Made in US
47-28-13-511 REV:A
CPDA-1
anticoagulant citrate phosphate dextrose adenine solution |
|||||||||||||||||||||
|
|||||||||||||||||||||
|
|||||||||||||||||||||
|
|||||||||||||||||||||
|
|||||||||||||||||||||
|
Labeler - Fenwal, Inc. (794519020) |