OXYGEN - oxygen gas 
Linde Gas & Equipment Inc.

Disclaimer: This drug has not been found by FDA to be safe and effective, and this labeling has not been approved by FDA. For further information about unapproved drugs, click here.

----------

OXYGEN, COMPRESSED

WARNINGS SECTION

WARNING: For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only.

Uninterrupted use of high concentrations of oxygen over a long duration, without monitoring its effect on oxygen content of arterial blood, may be harmful. Do not attempt to use on patients who have stopped breathing unless used in conjunction with resuscitative equipment. Patient monitoring is recommended for patient receiving supplemental oxygen. Flow rates for this product are approximate. Federal law restricts this device to sale by or on the order of a physician.

Keep out of reach and sight of children.

Federal law requires that this container be refilled with Oxygen USP only by establishments registered as a drug manufacturer in accordance with the Federal Food, Drug, and Cosmetic Act.

Note: Keep valve minimum 12 inches (30cm) away from magnet opening. Do not move the package during MRI scan, as the MR data quality may be degraded.

Un-Coded Section

MR
MR CONDITIONAL
Static Field Strength 3.0T
Static Field Gradient 375 Gauss/cm

Instructions for use
-Operating temperature -4°F to 113°F (-20°C to +45°C)
-Only use with devices intended for the product administration.
-Prepare to change the cylinder when the pointer on the gauge enters the red area.
-to ensure the correct, prescribed flow rate, always make sure that the flow selector is positioned on a numbered flow setting, not between flow settings.


1) Pressure outlet (if provided)
2) Shut-off valve
3) Gauge
4) Flow outlet
5) Flow selector

Before Use:
>Check gauge (3) for content.
>Check flow selector (5) is set at zero.
>Connect equipment to outlet (1 or 4).
>Slowly open shut off valve (2) fully by turning counterclockwise.
>If equipment is connected to the flow outlet (4) select flow by turning flow selector (5) stepwise up through flow settings.
The patient can now be connected to a delivery device.

After Use:
>Turn flow selector (5) to zero by turning the flow selector stepwise down through flow settings.
>Close shut off valve (2) by turning clockwise.
>Vent downstream equipment.
>Disconnect the equipment from the outlet (1 or 4).
 

Batch:


DO NOT REMOVE THIS LABEL


LG015C

 

PACKAGE LABEL.PRINCIPAL DISPLAY PANEL     

O2 Compressed Wall Label

Linde
OXYGEN, COMPRESSED USP UN1072

Produced By Air Liquefaction

CAS 7782-44-7

Distributed by:
Linde Gas North America LLC
200 Somerset Corporate Blvd., Suite 7000
Bridgewater, NJ 08807
866.543.3427

OXYGEN 2

MEDICAL GAS

WARNING: For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only.

Uninterrupted use of high concentrations of oxygen over a long duration, without monitoring its effects on oxygen content of arterial blood, may be harmful. Do not attempt to use on patients who have stopped breathing unless used in conjunction with resuscitative equipment. Federal law requires that this container be refilled with Oxygen USP only by establishments registered as a drug manufacturer in accordance with the Federal Food, Drug, and Cosmetic Act.

DANGER:
KEEP OUT OF REACH OF CHILDREN.
MAY CAUSE OR INTENSIFY FIRE: OXIDIZER
CONTAINS GAS UNDER PRESSURE: MAY EXPLODE IF HEATED.

Do not handle until all safety precautions have been read and understood. Keep and store away from clothing and other combustible materials. Keep valves and fittings free from grease and oil. Use and store only outdoors or in a well-ventilated place. In case of fire: Stop leak if safe to do so. Use a back flow preventive device in the piping. Use only with equipment of compatible materials of construction and rated for cylinder pressure. Use only with equipment cleaned for oxygen service. Open valve slowly. Close valve after each use and when empty. Protect from sunlight when ambient temperature exceeds 52°C (125°F). Read and follow the Safety Data Sheet (SDS) before use.

DO NOT REMOVE THIS LABEL.

CONTENTS:
Cylinder
Type
Liters  
T8584[_]
H6322[_]
K6322[_]
M3172[_]
F1567[_]
E680[_]
DD588[_]
D413[_]
C248[_]
BB147[_]
B147[_]
AA103[_]
AAA39[_]
M-7196[_]

LG010G

O2 Compressed Shoulder Label

Linde
OXYGEN, COMPRESSED USP

UN1072

Produced by Air Liquefaction

CAS 7782-44-7

DO NOT REMOVE THIS PRODUCT LABEL.

OXYGEN 2
LG011G

MEDICAL GAS

WARNING: For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Uninterrupted use of high concentrations of oxygen over a long duration, without monitoring its effects on oxygen content or arterial blood, may be harmful. Do not attempt to use on patients who have stopped breathing unless used in conjunction with resuscitative equipment.

DANGER: MAY CAUSE OR INTENSIFY FIRE; OXIDIZER. CONTAINS GAS UNDER PRESSURE; MAY EXPLODE IF HEATED. Do not handle until all safety precautions have been read and understood. Keep and store away from clothing and other combustible materials. Keep valves and fittings free from grease and oil. Use an store only outdoors or in a well-ventilated place. In case of fire: Stop leak if safe to do so. Use a back flow preventive device in the piping. Use only with equipment of compatible materials of construction and rated for cylinder pressure. Use only with equipment cleaned for oxygen service. Open valve slowly. Close valve after each use and when empty. Protect from sunlight when ambient temperature exceeds 52°C (125°F). Read and follow the Safety Data Sheet (SDS) before use.  

Distributed by: 
Linde Gas North America LLC
200 Somerset Corporate Blvd., Suite 7000
Bridgewater, NJ 08807
866.543.3427

CONTENTS:
Cylinder
Type
 Liters
T[_]8584
H[_]6322
K[_]6322
M[_]3172
____ ________
O2 Compressed Wall Label MRI Conditional

MR CONDITIONAL
Static Field Strength ≤ 3.0T
Spatial Gradient Field ≤ 100 gauss/cm

OXYGEN, COMPRESSED
USP
UN1072
CAS 7782-44-7
Produced by Air Liquefaction.

Distributed by:
Linde Gas North America, LLC
200 Somerset Corporate Blvd., Suite 7000
Bridgewater, NJ 08807
866.513.3427

CONTENTS:
Cylinder Type LitersNDC
E[_]68025373-001-10
D[_]41325373-001-01

No user-serviceable parts. Return to distributor for service/repair. MR conditional void if user-serviced.

No user transfill. Return to distributer for transfill.

OXYGEN 2

Linde

ACCURACY:±20%: 1, 1.5, 2 lpm
 ±15%: 3 lpm
 ±10%: 4 ,5 ,8 ,10 ,15 lpm
FOR MRI:derate by -3.5% at 1.5T
 derate by -12% at 3.0T

MR CONDITIONAL
Static Field Strength ≤ 3.0T
Spatial Gradient Field ≤ 100 gauss/cm

INDICATIONS FOR USE: Intended to provide supplemental oxygen to adults, MRI safe, MRI-compatible, and intended for use during MRI imaging for MRI systems up to 3.0T. For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only.

WARNING: For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Uninterrupted use of high concentrations of oxygen over a long duration, without monitoring its effect on oxygen content of arterial blood, may be harmful. Do not attempt to use on patients who have stopped breathing unless used in conjunction with resuscitative equipment. Keep out of reach of children. Federal law requires that this container be refilled with Oxygen USP only by establishments registered as a drug manufacturer in accordance with the Federal Food, Drug, and Cosmetic Act. DANGER: MAY CAUSE OR INTENSIFY FIRE; OXIDIZER. CONTAINS GAS UNDER PRESSURE; MAY EXPLODE IF HEATED. Do not handle until all safety precautions have been read and understood. Keep and store away from clothing and other combustible materials. Keep valves and fittings free from grease and oil. Use and store only outdoors or in a well-ventilated place. In case of fire: Stop leak if safe to do so. Use a back flow preventive device in the piping. Use only with equipment of compatible materials of construction and rated for cylinder pressure. Use only with equipment cleaned for oxygen service. Open valve slowly. Close valve after each use and when empty. Protect from sunlight when the ambient temperature exceeds 52°C (125°F). Read and follow the Safety Data Sheet (SDS) before use.

DO NOT REMOVE THIS LABEL.

MRI
CAUTION
≤ 3.0T
≤ 100 G/cm

NOTICE
-Open valve slowly. Close after each use.


-Environmental specifications for handling, storage, transport, and operating conditions: -20 to 125°F (-29 to 52°C).

LG013G

LIV MR Conditional Wall Label

THE LINDE GROUP
Linde

OXYGEN,
COMPRESSED
USP

OXYGEN
2

UN1072
CAS: 7782-44-7

Produced by Air Liquefaction
Oxygen (O2) 100%
under 2000 psi pressure (59°F or 15°C)

CONTENTS:
Cylinder type:LitersNDC
E[_]68025373-001-10
D[_]41325373-001-01

Service and transfill:
No user-serviceable parts. Return to distributor for service/repair. MRI safe and MRI compatible void if user-serviced. No user transfill. Return to distributor for transfill.


Distributed by:
Linde Gas North America LLC
200 Somerset Corporate Blvd., Suite 7000
Bridgewater, NJ 08807
866.543.3427

MR
MR CONDITIONAL


Static Field Strength 3.0T
Static Field Gradient 375 Gauss/cm

LIV® Linde Integrated Valve Portable oxygen system.


INDICATIONS FOR USE:
The LIV® is an integrated portable oxygen delivery system intended to provide supplemental oxygen to pediatrics and adults. The device is MR-conditional (per ASTM standard 2503-05), and intended for use during MR imaging for MRI systems up to 3.0T.


Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling or rolling.

DANGER: MAY CAUSE OR INTENSIFY FIRE; OXIDIZER. CONTAINS GAS UNDER PRESSURE; MAY EXPLODE IF HEATED.

Do not handle until all safety precautions have been read and understood. Keep and store away from clothing and other combustible materials. Keep valves and fittings free from grease and oil. Use and store only outdoors or in a well-ventilated place. In case of fire: Stop leak if safe to do so. Use a back flow preventive device in the piping. Use only with equipment of compatible materials of construction and rated for cylinder pressure. Use only with equipment cleaned for oxygen service. Open valve slowly. Close valve after each use and when empty. Protect from sunlight when ambient temperature exceeds 52°C (125°F). Read and follow the Safety Data Sheet (SDS) before use.

PACKAGE LABEL.PRINCIPAL DISPLAY PANEL

LINDE


MEDICAL GAS

O2 Compressed Wall Label LIV-Lg Pkg

THE LINDE GROUP
Linde

OXYGEN,
COMPRESSED
USP

OXYGEN
2

UN1072
CAS: 7782-44-7

Produced by Air Liquefaction
Oxygen (O2) 100%
under 3000 psi pressure (59°F or 15°C)

CONTENTS:
Cylinder TypeLitersNDC
E[_]100425373-001-15
D[_]61825373-001-16

Service and transfill:
No user-serviceable parts. Return to distributor for service/repair. MRI safe and MRI compatible void if user-serviced. No user transfill. Return to distributor for transfill.


Distributed by:
Linde Gas North America LLC
200 Somerset Corporate Blvd., Suite 7000
Bridgewater, NJ 08807
866.543.3427

MR
MR CONDITIONAL


Static Field Strength 3.0T
Static Field Gradient 375 Gauss/cm

LIV®

INDICATIONS FOR USE:
The LIV® is an integrated portable oxygen delivery system intended to provide supplemental oxygen to pediatrics and adults. The device is MR-conditional (per ASTM standard 2503-05), and intended for use during MR imaging for MRI systems up to 3.0T.


Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling or rolling.

DANGER: MAY CAUSE OR INTENSIFY FIRE; OXIDIZER. CONTAINS GAS UNDER PRESSURE; MAY EXPLODE IF HEATED.

Do not handle until all safety precautions have been read and understood. Keep and store away from clothing and other combustible materials. Keep valves and fittings free from grease and oil. Use and store only outdoors or in a well-ventilated place. In case of fire: Stop leak if safe to do so. Use a back flow preventive device in the piping. Use only with equipment of compatible materials of construction and rated for cylinder pressure. Use only with equipment cleaned for oxygen service. Open valve slowly. Close valve after each use and when empty. Protect from sunlight when ambient temperature exceeds 52°C (125°F). Read and follow the Safety Data Sheet (SDS) before use.

WARNING: For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only.

Uninterrupted use of high concentrations of oxygen over a long duration, without monitoring its effect on oxygen content of arterial blood, may be harmful. Do not attempt to use on patients who have stopped breathing unless used in conjunction with resuscitative equipment. Patient monitoring is recommended for patient receiving supplemental oxygen. Flow rates for this product are approximate. Federal law restricts this device to sale by or on the order of a physician.

Keep out of reach and sight of children.

Federal law requires that this container be refilled with Oxygen USP only by establishments registered as a drug manufacturer in accordance with the Federal Food, Drug, and Cosmetic Act.

Note: Keep valve minimum 12 inches (30cm) away from magnet opening. Do not move the package during MRI scan, as the MR data quality may be degraded.

DO NOT REMOVE THIS LABEL

MR
MR CONDITIONAL
Static Field Strength 3.0T
Static Field Gradient 375 Gauss/cm

Instructions for use
-Operating temperature -4°F to 113°F (-20°C to +45°C)
-Only use with devices intended for the product administration.
-Prepare to change the cylinder when the pointer on the gauge enters the red area.
-to ensure the correct, prescribed flow rate, always make sure that the flow selector is positioned on a numbered flow setting, not between flow settings.


1) Pressure outlet (if provided)
2) Shut-off valve
3) Gauge
4) Flow outlet
5) Flow selector

Before Use:
>Check gauge (3) for content.
>Check flow selector (5) is set at zero.
>Connect equipment to outlet (1 or 4).
>Slowly open shut off valve (2) fully by turning counterclockwise.
>If equipment is connected to the flow outlet (4) select flow by turning flow selector (5) stepwise up through flow settings.
The patient can now be connected to a delivery device.

After Use:
>Turn flow selector (5) to zero by turning the flow selector stepwise down through flow settings.
>Close shut off valve (2) by turning clockwise.
>Vent downstream equipment.
>Disconnect the equipment from the outlet (1 or 4).
 

Batch:


LG016B

O2 Compressed Wall Label LIV IQ

THE LINDE GROUP
Linde

OXYGEN,
COMPRESSED
USP

OXYGEN
2

UN1072
CAS: 7782-44-7

Produced by Air Liquefaction
Oxygen (O2) 100%
under 2000 psi pressure (59°F or 15°C)

CONTENTS:
Cylinder type:LitersNDC
E[_]68025373-001-10
D[_]41325373-001-01

Service and transfill:
No user-serviceable parts. Return to distributor for service/repair. MRI safe and MRI compatible void if user-serviced. No user transfill. Return to distributor for transfill.
Federal law requires that this container be refilled with oxygen USP only by establishments registered as a drug manufacturer in accordance with the Federal Food, Drug, and Cosmetic Act.

DO NOT REMOVE THIS LABEL


Distributed by:
Linde Gas North America LLC
200 Somerset Corporate Blvd., Suite 7000
Bridgewater, NJ 08807
866.543.3427

Note: Keep valve minimum 12 inches (30 cm) away from magnet opening. Do not move the package during MRI scan, as the MR data quality may be degraded.


MR
MR CONDITIONAL
Static Field Strength 3.0T
Static Field Gradient 375 Gauss/cm

LIV®IQ Linde Integrated Valve Portable oxygen system

INDICATIONS FOR USE:
The LIV®IQ is an integrated portable oxygen delivery system intended to provide supplemental oxygen to pediatrics and adults. The device is MR-conditional (per ASTM standard 2503-05), and intended for use during MR imaging for MRI systems up to 3.0T.


Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling or rolling.

DANGER: MAY CAUSE OR INTENSIFY FIRE; OXIDIZER. CONTAINS GAS UNDER PRESSURE; MAY EXPLODE IF HEATED.

Do not handle until all safety precautions have been read and understood. Keep and store away from clothing and other combustible materials. Keep valves and fittings free from grease and oil. Use and store only outdoors or in a well-ventilated place. In case of fire: Stop leak if safe to do so. Use a back flow preventive device in the piping. Use only with equipment of compatible materials of construction and rated for cylinder pressure. Use only with equipment cleaned for oxygen service. Open valve slowly. Close valve after each use and when empty. Protect from sunlight when ambient temperature exceeds 52°C (125°F). Read and follow the Safety Data Sheet (SDS) before use.

WARNING: For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only.

Uninterrupted use of high concentrations of oxygen over a long duration, without monitoring its effect on oxygen content of arterial blood, may be harmful. Do not attempt to use on patients who have stopped breathing unless used in conjunction with resuscitative equipment. Patient monitoring is recommended for patient receiving supplemental oxygen. Flow rates for this product are approximate. Federal law restricts this device to sale by or on the order of a physician.

Keep out of reach and sight of children.

Download the full user guide at: www.liv-iq.com or scan with your smart phone:



Guide to Digital Display
Selected Flow
Flow is active
New, full cylinder
Therapy time remaining at selected flow setting
Remaining gas volume, shown 3/4 full
Mute button

If there are no lines or numbers around the content indicator, do not use the product.

MR
MR CONDITIONAL
Static Field Strength 3.0T
Static Field Gradient 375 Gauss/cm

Instructions for use
-Operating temperature -4°F to 113°F (-20°C to +45°C)
-Only use with devices intended for the product administration.
-Prepare to change the cylinder when the pointer on the gauge enters the red area.
-to ensure the correct, prescribed flow rate, always make sure that the flow selector is positioned on a numbered flow setting, not between flow settings.


1) Pressure outlet (if provided)
2) Shut-off valve
3) Flow selector
4) Flow outlet
5) Digital display
Mute button

Before Use:
>Check digital display (5) for content.
>Check flow selector (3) is set at zero.
>Connect equipment to outlet (1 or 4).
>Slowly open shut off valve (2) slowly to "ON".
>Turn the flow selector (3) to the desired flow.

The patient can now be connected to a delivery device.

After Use:
>Remove the delivery device from the patient.
>Turn flow selector (3) to zero.
>Turn the shut off valve (2) to "OFF".
>Turn the flow selector (3) to a high flow to vent.
>Turn the flow selector (3) back to zero.
>Disconnect the delivery device from outlet (1 or 4).

-Using both outlets at the same time is not recommended and may cause a "HI FLO" alarm. To silence the alarm and continue, press and hold the mute button for 3 seconds.
-The mute button can silence any alarm for 1 minute (short press) or permanently (press & hold for 3 seconds).
 

Set flow selector to zero before opening shut-off valve.

Blocked tube/Low flow


Has the Mute button been pressed?
Yes: An active alarm has been muted.
No: Due to a rapid change in temperature, the low flow or high flow alert is temporarily disabled.


Batch:



LGIQ015B

OXYGEN  
oxygen gas
Product Information
Product TypeHUMAN PRESCRIPTION DRUGItem Code (Source)NDC:25373-001
Route of AdministrationRESPIRATORY (INHALATION)
Active Ingredient/Active Moiety
Ingredient NameBasis of StrengthStrength
Oxygen (UNII: S88TT14065) (Oxygen - UNII:S88TT14065) Oxygen99 L  in 100 L
Packaging
#Item CodePackage DescriptionMarketing Start DateMarketing End Date
1NDC:25373-001-01413 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
2NDC:25373-001-02640 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
3NDC:25373-001-036452 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
4NDC:25373-001-0434 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
5NDC:25373-001-05113 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
6NDC:25373-001-06170 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
7NDC:25373-001-07170 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
8NDC:25373-001-08196 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
9NDC:25373-001-09248 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
10NDC:25373-001-10680 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
11NDC:25373-001-171567 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
12NDC:25373-001-183172 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
13NDC:25373-001-196322 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
14NDC:25373-001-208584 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
15NDC:25373-001-151004 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
16NDC:25373-001-16618 L in 1 CYLINDER; Type 0: Not a Combination Product11/19/1990
Marketing Information
Marketing CategoryApplication Number or Monograph CitationMarketing Start DateMarketing End Date
unapproved medical gas11/19/1990
Labeler - Linde Gas & Equipment Inc. (805568339)
Registrant - Linde Gas & Equipment Inc. (805568339)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.010263301manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.016271697manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.833132561manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.063018874manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.069239874manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.784579190manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.016300933manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.164139060manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.026921048manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.129012352manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.142591424manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.168248081manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.067906953manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.789082976manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.149324605manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.833120244manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.833120723manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.134971428manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.839176356manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.069232473manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.160443631manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.833132108manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.080006834manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.080428082manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.019551467manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.078692645manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.181444840manufacture(25373-001)
Establishment
NameAddressID/FEIBusiness Operations
Linde Gas & Equipment Inc.805823411manufacture(25373-001)

Revised: 12/2021
Document Id: f65fbfa0-f21d-4b9a-98e0-5939f7c5456b
Set id: d41dee2b-8191-a5ec-6ca8-5b8edd7e3009
Version: 15
Effective Time: 20211214
 
Linde Gas & Equipment Inc.