CPDA-1 - anticoagulant citrate phosphate dextrose adenine solution
Fenwal, Inc.
----------
Contains Fenwal Express System and Y-Sampling Site for the collection of unanticoagulated whole blood samples for laboratory testing.
Integral filter unit intended for leukocyte reduction of Whole Blood up to 8 hours after blood collection when Whole Blood is stored at ambient temperature or up to 72 hours after blood collection when Whole Blood is refrigerated. The leukocyte reduced blood products may then be stored for the maximum allowable dating period.
Instructions for Use
Collection Procedure:
Use aseptic technique.
Precautions:
1. Identify Blood-Pack unit using appropriate donor identification system.
3. Clamp donor tubing between HighFlo1 needle and primary container. This step may be performed prior to step 1 or 2.
Precaution: Do not proceed with the remaining steps until the entire whole blood unit is collected.
11. To avoid possible contamination of the whole blood unit, before filling whole blood sample tubes, hermetically seal the donor tubing near the Y-Sampling Site on the side leading to the primary container using a metal clip or appropriate alternate method.
Precaution: Complete steps 12 - 21 within approximately 4 minutes after sealing the donor tubing to avoid possible clot formation in the tubing.
12. To collect samples, insert the access device by pushing firmly into the Y-Sampling Site until the membrane seal is penetrated (see Figure 2).
13. Open the cap on the access device (if applicable).
14. Directly align the vacuum sample tube with the internal needle in the access device. Insert vacuum sample tube into device until the stopper is punctured.
15. Allow vacuum sample tube to fill with blood then remove from the access device.
16. Repeat steps 14 and 15 until the desired number of vacuum sample tubes have been filled.
Precaution: When replacing the access device, be careful to avoid contact with any blood droplets on the Luer or Sampling Site. Discard used access device appropriately.
17. Release remaining pressure on donor’s arm.
18. If appropriate, apply clamp to donor tubing between needle and Y-Sampling Site.
19. Withdrawal of Needle (see Figure 3)
Precaution: The needle guard must be held stationary while the needle is withdrawn into it.
20. Remove and discard the Y-Sampling Site and needle in needle guard into an appropriate biohazardous waste container following established procedures. If donor tubing is also to be discarded, hermetically seal donor tubing directly above the primary container and remove.
Note: Step 21 may be performed prior to step 20 if desired.
21. If the donor tubing is not hermetically sealed directly above the primary container, then strip the blood from the remaining donor tubing into the primary container. Mix and allow tubing to refill; repeat once.
Filtration Procedure:
Precaution: Whole blood collected from certain donors may have extended filtration times and the potential for ineffective filtration and leukoreduction.
22. Mix unfiltered Whole Blood thoroughly. Invert the primary container and hang the filter set such that the filter remains vertical. To achieve maximum flow rate, allow set to hang to full length.
Note: The filtered Whole Blood container must remain below the level of the filter during filtration. For proper air expression to occur, ensure the filtered Whole Blood container is vertical.
23. Inspect all tubing to insure it hangs freely without kinks. Install and close clamp on bypass line.
24. Break the in-line cannula above the filter to start filtration. To completely break the in-line cannula, grasp with both hands. Snap it at a 90° angle in one direction, and then bend it at a 90° angle in the opposite direction. Allow filtration to continue until flow stops.
Note: Manual or mechanical pressure should not be used to increase the flow rate through the filter.
25. Open the clamp on the bypass line and allow air to transfer from the filtered Whole Blood container to the primary container.
Note: If desired, gently squeeze the filtered Whole Blood container to transfer remaining air through the bypass line.
26. Allow filtration to continue until the inlet side of the filter is filled with air.
27. Hermetically seal and separate the bypass line above the filtered Whole Blood container. Also hermetically seal and separate the segment line tubing directly above the top donor segment number. Use care to avoid fluid splatter. Discard filter and primary container appropriately.
Note: If a QC sample is desired, thoroughly mix the filtered Whole Blood and strip the donor segment tubing or bypass line tubing as desired. Use the tubing farthest away from the filtered Whole Blood container as the QC sample.
28. Make donor segments. Leave segments attached to the filtered Whole Blood container.
29. At the appropriate time, prepare the Blood-Pack unit for centrifugation by thoroughly mixing the primary container end over end, then load the unit in a centrifuge cup per the instructions on page 3.
30. Following centrifugation, remove containers from the centrifugation cup taking care not to disturb the red blood cell / plasma interface.
31. Place filtered Whole Blood container in plasma extractor, and express plasma into empty Transfer-Pack container by releasing pressure plate and opening closure in tubing of filtered Whole Blood container.
32. When the desired amount of plasma has been removed, clamp the tubing between Y and plasma container. Hermetically seal and separate transfer tubing being careful to avoid fluid splatter.
33. Hermetically seal and separate transfer tubing near the container of filtered Red Blood Cells. Be careful to avoid fluid splatter.
34. For further processing of plasma product with multiple Blood-Pack units, use standard component processing and storage techniques.
Note: Fresh Frozen Plasma should be separated from the red blood cells and placed in the freezer at -18°C or colder within 8 hours after blood collection.
35. Store suspended CPDA-1 Whole Blood/Red Blood Cells, Leukocytes Reduced, (as applicable) between 1 and 6°C.
36. Infuse CPDA-1 Whole Blood/Red Blood Cells, Leukocytes Reduced, within 35 days of collection.
Warning: Failure to achieve closed system processing conditions negates the extended storage claim and the red blood cell product must be transfused within 24 hours.
Store at Controlled Room Temperature. Protect from freezing. Avoid excessive heat. Definition of “Controlled Room Temperature”: “A temperature maintained thermostatically that encompasses the usual and customary working environment of 20° to 25°C (68° to 77°F); that results in a mean kinetic temperature calculated to be not more than 25°C; and that allows for excursions between 15°C and 30°C (59° and 86°F) that are experienced in pharmacies, hospitals, and warehouses. Provided the mean kinetic temperature remains in the allowed range, transient spikes up to 40°C are permitted as long as they do not exceed 24 hours ...The mean kinetic temperature is a calculated value that may be used as an isothermal storage temperature that simulates the non isothermal effects of storage temperature variations.”
Reference: United States Pharmacopeia, General Notices. United States Pharmacopeial Convention, Inc. 12601 Twinbrook Parkway, Rockville, MD.
1 Van der Meer, P.F., & de Korte, D. “Increase of blood donation speed by optimizing the needle-to-tubing connection: an application of donation software.” Vox Sanguinis 2009, 97: 21-25
Sorvall is a trademark of Thermo Fisher Scientific LLC. Sepacell is a trademark of Asahi Kasei Medical Co., Ltd.
© 2019 Fresenius Kabi AG. All rights reserved
Manufacturer
Fresenius Kabi AG
61346 Bad Homburg / Germany
www.fresenius-kabi.com
1-800-933-6925
47-23-13-838 REV: A
Code 4R3313E 12 Units
Fresenius Kabi
Fenwal Blood-Pack Units Double
Anticoagulant Citrate Phosphate Dextrose Adenine Solution, USP (CPDA-1) Blood-Pack Unit; Integral Sepacell RS-2000 Whole Blood Leukocyte Reduction Filter
For the Collection of 500 mL Blood
Fenwal Express System, Y-Sampling Site, 16 ga. Ultra Thin Wall Fenwal HighFlo Needle
Rx only
Each unit consists of a primary container with 70 mL of CPDA-1 solution containing 2.23 g Dextrose (monohydrate) USP, 1.84 g Sodium Citrate (dihydrate) USP, 209 mg Citric Acid (anhydrous) USP, 155 mg Monobasic Sodium Phosphate (monohydrate) USP and 19.3 mg Adenine USP, pH may have been adjusted with sodium hydroxide; an integral RS-2000 filter with one empty 450 mL PL 146 Plastic container for red cell storage and one empty 400 mL Transfer Pack container.
Sterile, non-pyrogenic fluid path.
See instructions for use.
Single use only.
Store at Controlled Room Temperature (refer to direction insert).
Sepacell is a trademark of Asahi Kasei Medical Co., Ltd.
Manufacturer
Fresenius Kabi AG
61346 Bad Homburg / Germany
www.fresenius-kabi.com
Made in USA
47-28-13-840 REV: A
CPDA-1
anticoagulant citrate phosphate dextrose adenine solution |
|||||||||||||||||||||
|
|||||||||||||||||||||
|
|||||||||||||||||||||
|
|||||||||||||||||||||
|
|||||||||||||||||||||
|
Labeler - Fenwal, Inc. (794519020) |
Establishment | |||
Name | Address | ID/FEI | Business Operations |
---|---|---|---|
Fenwal International, Inc. | 091164590 | MANUFACTURE |