CPDA-1 - anticoagulant citrate phosphate dextrose adenine solution
Fenwal, Inc.
----------
Contains Sample Diversion System for the collection of unanticoagulated whole blood samples for laboratory testing.
The elongated primary container may be used to collect and store 450 mL of blood for 35 (CPDA-1) days. It may also be used to prepare and store frozen red cells using the high glycerol (40% W/V) technique.
Instructions for Use
Collection Procedure:
Use aseptic technique.
Notes:
Precautions:
1. Identify Blood-Pack unit using appropriate donor identification system.
2. Adjust donor scale to desired collection weight.
3. Position primary container on the donor scale as far as possible below donor arm.
4. Clamp donor tubing between the needle and Y-junction. (This step can be performed prior to step 1 or 2.)
5. Visually inspect the tubing from the in-line cannula to the sample pouch and donor needle, as well as the sample pouch to reconfirm that there is no anticoagulant present.
Note: Ensure that the sample pouch remains below the donor’s arm.
6. Following blood center procedures, apply pressure to donor’s arm and disinfect site of venipuncture.
7. Remove needle cover per instructions below:
8. Following blood center procedures, perform venipuncture, appropriately secure donor needle and/or tubing and release clamp on donor tubing.
9. When good blood flow is established, stabilize the front of the needle guard to arm with tape (see Figure 2).
10. Allow the sample pouch to fill with blood according to center procedure. Monitor blood flow into sample pouch.
Notes:
Precautions:
11. Close the blue clamp on tubing between the Y-junction and the sample pouch.
12. Break the in-line cannula below the Y-junction in the donor tubing to the primary container allowing blood collection to proceed. To completely break the in-line cannula, grasp with both hands. Snap it at a 90° angle in one direction, and then bend it at a 90° angle in the opposite direction. Ensure the in-line cannula is completely broken and that the blood flows freely to the primary container.
Precaution: Failure to break the in-line cannula completely may result in restricted blood flow.
13. Mix blood and anticoagulant in primary container immediately, at several intervals during collection and immediately after collection.
14. Following blood center procedures, hermetically seal the tubing between the sampling site and the Y-junction to maintain sterility of the blood collection system prior to removing blood samples.
Warning:
• Do not proceed with the remaining steps until the tubing leading to the sample pouch is hermetically sealed between the sampling site and the Y-junction. To maintain the whole blood collection container as a closed system, the tubing between the sample pouch and Y-junction must be hermetically sealed prior to inserting the access device into the sampling site. Failure to do so may lead to contamination of the whole blood collection.
15. To collect samples, insert the access device by pushing firmly into the sampling site until the membrane seal is penetrated.
Note: If the access device is assembled such that the outer barrel is screwed onto the Luer, make sure to rotate clockwise upon insertion to avoid barrel detaching from Luer.
16. Open the cap on the access device (if applicable). Hold access device so that the sample pouch hangs down.
17. Directly align the vacuum sample tube with the internal needle in the access device. Insert vacuum sample tube into device.
18. Allow vacuum sample tube to fill with blood then remove from the access device.
19. Repeat steps 17 and 18 until the desired number of vacuum sample tubes have been filled.
Notes:
20. Collect only 450 ± 45 mL of blood into the elongated primary container.
Precaution: Once the desired blood volume is collected, complete steps 21–24 within approximately 4 minutes to avoid possible clot formation in the tubing.
21. Release pressure on the donor’s arm. If appropriate, apply clamp to donor tubing between the needle and the Y-junction.
22. Hermetically seal donor tubing near in-line cannula on side leading to the primary container.
23. Withdrawal of Needle (see Figure 3)
Precaution: The needle guard must be held stationary while the needle is withdrawn into it.
24. Strip blood from donor tubing into primary container, mix and allow the tubing to refill; repeat once.
25. Seal at X marks on donor tubing to provide numbered aliquots of anticoagulated blood for typing or crossmatching.
Note: Step 26 may be performed prior to step 24 or 25 if desired.
26. Remove and discard the Sample Diversion System and needle in the needle guard into an appropriate biohazardous waste container following established procedures.
27. For component preparation, prior to centrifugation, the excess material at bottom of the elongated primary container should be rolled over as follows:
28. At the appropriate time, prepare the Blood-Pack unit for centrifugation by thoroughly mixing the primary container end over end, then load the unit in a centrifuge cup per the instructions on page 4.
29. Gently remove the (bottom-rolled) elongated primary container from centrifuge cup. Using care to ensure that bottom fold area is supported, place the container in the plasma extractor.
30. Express plasma into empty Transfer Pack container by releasing pressure plate and opening closure in tubing of primary container. All visible (platelet rich) plasma should be removed if cells are to be frozen subsequently.
31. When desired amount of plasma has been removed, clamp tubing between Y and plasma container, remove tape, and unfold unit.
32. Seal transfer tubing in three places between the Y-connectors. Cut middle seal being careful to avoid fluid splatter.
33. For further processing with multiple Blood-Pack units, use standard component processing and storage techniques.
For Liquid Red Cell Storage:
34. Store suspended CPDA-1 Whole Blood/Red Blood Cells between 1 and 6°C and
infuse within 35 days of collection.
For Frozen Red Cell Storage:
Precaution: Freeze within 6 days of collection. Use aseptic technique during red cell glycerolization.
35. If remaining Red Blood Cells are intended for frozen storage in the elongated primary container, the hematocrit should be between 75 and 80% and one empty container should be left attached to the primary for glycerol removal.
36. Supplies
37. Allow glycerol solution and red cells to warm to 22–30°C prior to adding solution to red cells. Process cells within 4 hours after removal from refrigeration.
38. Remove all but one numbered tubing segment. Heat seal the remaining segment as close to the container as possible. Leave the segment attached to the container.
39. Weigh the unit of red cell concentrate and subtract the tare weight (approx. 72 grams for the elongated primary container and integrally attached Transfer Pack container). The net weight of the red cells should not exceed 330 grams.
40. Place the red cell concentrate on a platform agitator.
41. Remove the metal tab from the top of the glycerol solution bottle, disinfect the stopper with 70% isopropyl alcohol and aseptically insert one of the plasma transfer set couplers into the bottle outlet port. Close roller clamp on plasma transfer set.
42. Insert the filtered airway needle into the bottle vent port.
43. Invert the bottle and suspend it 18 inches above the level of the primary bag on the agitator.
44. Aseptically connect the remaining transfer set connector to the delivery tube leading from the elongated primary container.
45. Glycerol solution should be added to cells based on cell volume and in three steps. Refer to Table 1.
46. Adjust platform agitator to 180 oscillations/minute. Occlude tubing between elongated primary collection container and empty Transfer Pack container by folding tubing and placing a hand sealer clip over fold (do not crimp) to prevent flow of solution into Transfer Pack container.
47. Add 50 mL of glycerol solution to the red cells while the cells are being agitated.
48. Turn off agitator and allow equilibration for 5 minutes.
49. Restart agitator; add 50 mL of glycerol solution to cells.
50. Turn off agitator and allow equilibration for 2 minutes.
51. Remove red cells from agitator.
52. Refer to Table 1 to determine remaining volume of glycerol solution to be added to red cells. Use continuous, vigorous manual agitation during solution addition.
53. Seal tubing proximal to the delivery tube assembly. Ensure that the Transfer Pack container remains integrally attached to the primary collection bag. Discard the glycerol bottle, connector tubing, and delivery tube assembly.
54. Centrifuge the glycerolized red cells at 1250 x g for 10 minutes.
55. Suspend primary container in plasma extractor, remove hand sealer clip, and express all visible supernatant glycerol from the red cell container into the Transfer Pack container.
56. Seal the tubing 2 inches from the red cell container, separate tubing and discard the Transfer Pack container containing the supernatant glycerol solution.
57. Appropriately label product (Figure 5A).
58. Fold over the top portion of the primary bag approximately 2 inches (Figure 5B), place the unit into the 8" x 12" plastic overwrap (Figure 5C) and seal across the top using an impulse sealer so that there is as little trapped air as possible. Make sure that the ports and tubing segments are folded beneath the unit so that they are protected from breakage when frozen.
59. Place the glycerolized red cells and overwrap into the cardboard box (Figure 5D). Close the box and seal the flaps with tape. Label the box, and place in the bottom of a -80°C freezer. To facilitate freezing, do not stack units. After 24 hours, the
frozen units can be stacked and/or stored in other -80°C freezers.
60. After storage, thaw and deglycerolize red cells according to standard institutional procedure.
61. Infuse deglycerolized cells within 24 hours after processing.
References
Freezing in the Primary Polyvinylchloride Plastic Collection Bag: A new system for freezing nonrejuvenated and rejuvenated Red Blood Cells. C.R. Valeri, D.A. Valeri, J. Anastasi, J.J. Vecchione, R.C. Dennis, and C.P. Emerson. Transfusion 1981; 21:138-149.
Store at Controlled Room Temperature. Protect from freezing. Avoid excessive heat. Definition of “Controlled Room Temperature”:
“A temperature maintained thermostatically that encompasses the usual and customary working environment of 20° to 25°C (68° to 77°F); that results in a mean kinetic temperature calculated to be not more than 25°C; and that allows for excursions between 15°C and 30°C (59° and 86°F) that are experienced in pharmacies, hospitals, and warehouses. Provided the mean kinetic temperature remains in the allowed range, transient spikes up to 40°C are permitted as long as they do not exceed 24 hours ... The mean kinetic temperature is a calculated value that may be used as an isothermal storage temperature that simulates the non isothermal effects of storage temperature variations.”
Reference: United States Pharmacopeia, General Notices. United States Pharmacopeial
Convention, Inc. 12601 Twinbrook Parkway, Rockville, MD.
1Van der Meer, P.F., & de Korte, D. “Increase of blood donation speed by optimizing the needle-totubing connection: an application of donation software.” Vox Sanguinis 2009, 97: 21-25.
Sorvall is a trademark of Thermo Fisher Scientific LLC.
© 2019 Fresenius Kabi AG. All rights reserved.
Manufacturer
Fresenius Kabi AG
61346 Bad Homburg / Germany
www.fresenius-kabi.com
1-800-933-6925
47-23-13-769 REV: A
Code 4R1250 3 Units
Fenwal Blood-Pack Units Quadruple
Anticoagulant Citrate Phosphate Dextrose Adenine Solution, USP (CPDA-1) with Elongated Primary Container
For Collection of 450 mL Blood and Optional Freezing of Red Blood Cells in the Primary Container Sample Diversion System, 16 ga. Ultra Thin Wall Fenwal HighFlo Needle
Rx only
Each unit consists of an elongated primary container with delivery tube assembly and with 63 mL of CPDA-1 solution containing 2 g Dextrose (monohydrate) USP, 1.66 g Sodium Citrate (dihydrate) USP, 188 mg Citric Acid (anhydrous) USP, 140 mg Monobasic Sodium Phosphate (monohydrate) USP and 17.3 mg Adenine USP, pH may have been adjusted with sodium hydroxide; three empty 400 mL Plastic Transfer-Pack containers.
Sterile, non-pyrogenic fluid path
See instructions for use.
Single use only.
Store at Controlled Room Temperature (refer to direction insert). Protect from freezing. Avoid excessive heat.
Manufacturer
Fresenius Kabi AG
61346 Bad Homburg / Germany
www.fresenius-kabi.com
Made in US
47-28-13-770 REV: A